What is the Accuracy and Precision of the Microlab 600 ?

The Bubble Free Prime syringes are specifically designed for the Microlab 600 to enable the highest Accuracy and Precision. They are Gastight syringes with a unique plunger tip. The plunger tip extends through the top of the syringe and slightly into the valve. These syringes are designed to reduce dead volume, improve priming, and limit sample carryover.

Syringe Size Percent Stroke Accuracy (±) Precision
10 µL 5% ≤ Stroke < 30% 3.0% 2.0%
30% = Stroke 2.0% 0.2%
Stroke = 100% 1.0% 0.2%
25 µL 5% ≤ Stroke < 30% 3.0% 2.0%
30% = Stroke 1.5% 0.2%
Stroke = 100% 1.0% 0.2%
50 µL 5% ≤ Stroke < 30% 3.0% 1.5%
30% = Stroke 1.2% 0.5%
Stroke = 100% 1.0% 0.2%
100 µL 5% ≤ Stroke < 30% 3.0% 1.5%
30% = Stroke 1.2% 0.5%
Stroke = 100% 1.0% 0.2%
250 µL 5% ≤ Stroke < 30% 3.0% 1.5%
30% = Stroke 1.2% 0.5%
Stroke = 100% 1.0% 0.2%
500 µL 1% ≤ Stroke < 5% 3.0% 1.5%
5% ≤ Stroke < 30% 1.2% 0.5%
30% = Stroke 1.0% 0.2%
Stroke = 100% 1.0% 0.2%
1000 µL and larger 1% ≤ Stroke < 5% 3.0% 1.5%
5% ≤ Stroke < 30% 1.2% 0.5%
30% = Stroke 1.0% 0.2%
Stroke = 100% 1.0% 0.2%