ILM04.0/05.0 Method 200.7 Method 6010B Brief definition of function according to ILM04.0/05.0
Analytical Sample Sample - this method does
not clearly indicate counting
rules
Sample - this method does
not clearly indicate counting
rules
Everything other than the ICV, ICB, CCV, CCB, calibration standards and calibration
blank.
Sample Delivery Group
(SDG)
Group of 20 samples Not clearly stated A unit within a sample case that is used to identify a group of samples for delivery.
An SDG is a group of 20 or fewer real samples within a case. Note that since
sample deliveries will not include Preparation Blanks, ICSA and ICSAB and other
solutions that are counted as Analytical Samples, the number of Analytical
Samples for an SDG may be higher than 20.
Contract Required
Detection Limit (CRDL)
No equivalent Estimated IDLs are provided
but not mandated.
The detection limits for each of the 22 elements that must be met to comply
with the Statement of Work or Method.
Instrument Detection
Limit (IDL)
Method detection limits
(MDL) - 7 replicates of fortified
reagent water at 2-3 x
instrument detection limit
Method detection limits
(MDL) - at 3-5 times anticipated
detection limits - 7
replicates on 3 non-consecutive
days 'for additional
confirmation'.
The detection limit calculated by multiplying by 3 the average of the standard
deviations obtained on 3 nonconsecutive days from a standard solution at a
concentration of 3-5 times the instrument manufacturers suggested IDL.
Seven consecutive measurements are taken to define the standard deviation
under the same conditions as the proposed analytical method.
Initial Calibration
Verification (ICV)
See IPC below. ICV Used to initially verify the validity of the calibration by measuring the recovery of
analytes in this standard immediately after calibration. A second source standard.
Initial Calibration
Blank (ICB)
Calibration blank Calibration blank A “calibration blank” that immediately follows the ICV. Usually compared to
the Instrument Detection Limit (IDL).
Continuing Calibration
Verification (CCV)
Initial performance check
(IPC) standard - combines
both ICV and CCV - first
time limits are + 5%
CCV - accompanied by calibration
blank - can use ICV
instead - ± 10% limits plus
< 5 %RSD
Used to verify the validity of the calibration on an on-going basis. This standard
is measured at a frequency of every 10 'analytical samples'. Recovery
limits + 10%
Continuing Calibration
Blank (CCB)
Calibration blank Calibration blank A 'calibration blank' that immediately follows the CCV. Usually compared to
the Instrument Detection Limit (IDL).
Contract Required
Detection Limit Test for
ICP (CRI)
No equivalent No equivalent A standard containing the analytes at 2 times the CRDL or IDL whichever is
greater. No control limits yet applicable. Not required for Al, Ba, Ca, Fe, Mg,
Na and K. Analyzed at the beginning and end of the SDG.
Interference Check
Solution (ICSA)
Spectral interference check
(SIC) solutions - up to 17
interferent solutions tested -
test for 10% error in baseline
Interference check sample -
analytes at 0.5-1.0 mg/L -
interferents at 100 mg/L of
Al, Ca, Cr, Cu, Fe, Mg, Mn, Ni,
Ti and V are shown in Table 2
of the method. - test for 20%
error in baseline
Used to check the effect of interferents on the determination of the analytes.
ICSA is the interferents only solution and contains Al, Ca and Mg at 500
mg/L and Fe at 200 mg/L. Analytes should be within 0 ± 2 x CRDL. Analyzed
at the beginning and end of the SDG. Note the 20% error limit in the 6010B
method probably refers to ± 10% since otherwise the method is exactly the
same as 200.7.
Interference Check
Solution (ICSAB)
See above. See above. Used to check the effect of interferents on the determination of the analytes.
ICSAB contains both interferents and 16 selected analytes. Analytes must be
recovered within ± 20%. Analyzed at the beginning and end of the SDG.
Preparation Blank
(Prep Blk)
Lab reagent blank (LRB) Method blank Also known as a Reagent Blank. A volume of deionized distilled water
processed through the sample preparation procedure. Analytes are then monitored
versus the CRDL. In ILM04.0/05.0, no preparationblank correctionof
samples is done. One per SDG.
No equivalent Lab fortified blank (LFB) - an
aliquot of LRB which has
been spiked with analytes -
one per SDG
No equivalent  
Lab Control Sample
(LCS)
Quality control sample
(QCS) - 3 analyses to
within + 5% recovery
No equivalent A control standard sourced from the EPA or another independent source. The
LCS is put through the sample preparation process and then the recovery of
the analyes is calculated and compared to ± 20% limits. One per SDG.
Duplicate (Dup) Laboratory duplicates (LD1
and LD2)
Matrix spiked duplicate
Samples - measure 2 duplicates
of a spiked sample -
RPD to 20% - and spike
recovery to ± 25%
A duplicate aliquot of a sample, put through the sample preparation procedure.
Acts as a monitor for contamination and losses during sample preparation.
The Relative Percent Difference between the duplicate and the sample is
calculated and compared to ± 20% limits or CRDL limits. One per SDG. Note
the Matrix Spiked Duplicate of 6010B tries to combine both the Matrix Spike
and Duplicate tests into one but most ICP software is currently not designed
to handle this combination.
Serial Dilution (Ser) Dilution test Dilution test - do for new or
unusual matrices
Conduct a five fold dilution on one sample from each SDG. The calculated
result after correction for dilution must agree within ± 10% of the undiluted
sample result.
Spiked Sample
Analysis (SSA) or
Matrix Spike (MS)
Lab fortified matrix (LFM) -
spike every 10 % of samples
- recover to ± 30%
No equivalent (see PDSA) A spike is added to a sample prior to the digestion or sample preparation procedures.
The recovery of the spike is calculated and compared to ± 25% limits. A
matrix spike is required for each SDG.
Linear Range
Analysis (LRA)
Linear dynamic range
(LDR) - verified annually -
top standard recovered to -
10% limit - dilute all samples
that are more than
90% of the LDR
LDR - as per 200.7 A standard which is analyzed quarterly to confirm the linearity of analytical
calibrations. A high level standard must be recovered within 5% of the true
value. This defines the upper limit of the linear dynamic range.
Post Digestion Spike -
only required if the
SSA fails
Analyte addition test -
spike at 20-100 times the
MDL - recover to ± 15%
limits
Post digestion spike addition
(PDSA) - do for new or
unusual matrices - spike at
10-100 times the MDL -
recover to ± 25% limits
Post digestion spikes are often used to assess whether the Method of
Standard Additions is required. This is done separately from pre-digestion
spikes because the pre¬digestion spike might be indicative of contamination
picked up during the sample preparation procedure.

Back to EPA Standards

Shopping cart

Your cart is empty.

Return to shop
close