Intended use

The Microlab 600 is a highly precise syringe pump with a graphical user interface designed to quickly and easily dilute and dispense liquids.
In the dual syringe dispenser configuration, each syringe fills from a separate reservoir and dispenses separately from the hand probe.
This positive displacement system provides better than 99% accuracy, independent of a liquid's viscosity, vapor pressure and temperature. The inert fluid path minimizes sample carry over and improves compatibility with harsh chemicals.

The Hamilton Diluter Dispenser is shipped complete with Concorde Hand Probe, Universal Valves, Fill/Dispense Tubing, Accessory Holder & Country Specific Power.

  • Demonstrating dual syringe dispenser method ML600

    All Labs Can Use the Hamilton ML-600 Dual Syringe Dispenser

    Every laboratory has tasks too small to automate and too large to reliably accomplish by hand. The Microlab 600 is a semi-automated liquid handler designed specifically for these in-between applications that increase throughput and consistency while reducing cost and wasted buffer. Common industries using the Hamilton ML-600 Dual Syringe Dispenser are:
    ■ Pharmaceutical
    ■ Forensics
    ■ Environmental Analysis
    ■ Mining
    ■ Manufacturing

  • Technical specifications

    Specifications: MICROLAB 600 Syringe Pump
    Accuracy: +/- 1%
    Precision: +/- 0.2%
    Flow Rate: 0.003 - 6000 µL/second (depending on the syringe that is selected)
    Syringe Resolution: 0.02% of the nominal syringe volume
    Compatible Syringes; 10, 25, 50, 100, 250, 500 µL, 1, 2.5, 5, 10, 25, and 50 mL
    Fluid Path: Borosilicate, PTFE, CTFE
    Communication Type: Ethernet, 10/100 BASE-T
    Communication Protocol: .NET 2.0 Application Programming Interface (API)
    Pump Memory: One method stored in non volatile memory
    Calibration: Factory tested and traceable to N.I.S.T. standards
    Certifications: CE, CSA
    Power Rating: 24 VDC, 2.5A
    Dimensions: 7 x 5.5 x 10.5 inch (177.8 x 139.7 x 266.7 mm)
    Weight: 13 lbs (5.9 kg)
    Specifications: MICROLAB 600 Controller
    Screen Size: 5.7 inch (15cm diagonal)
    Screen Resolution: 640 (H) x 480 (V) pixels
    Tilt Positions: 5 positions from 90° to flat
    Mounting Options: On top of the Syringe Pump, Bench Top, or Wall Mount
    Program Memory: 2 GB (With Advanced Upgrade)
    Communication Type: Ethernet, 10/100 BASE-T
    Power Rating: 24VDC, 2.5A
    Dimensions: 2.25 x 6.5 x 7 inch (57.2 x 165.1 x 177.8mm) in down position
    Weight: 1.9 lbs (0.86 kg)


    An accuracy of 1% and 0.2% precision is achieved when dispensing above 10% of a syringes nominal volume. Reasonable accuracy can be achieved down to 1% of a syringes nominal volume. The Microlab 600 syringe allows for programming of dispense volumes in increments less than 0.01% of the nominal syringe volume.

  • How to set up and run the Microlab ML600 Diluter / Dispenser from unpacking to programming.

    This playlist of videos will show you how to set up your ML600 from unpacking to getting the instrument ready for use. The topics covered are;
    1. 1. Unpack and setup the ML600
    2. 2. General Information and tips for the ML600
    3. 3. Introduction to the Dilution Wizard
    4. 4. How to prevent carryover issues in dilutions
    5. 5. Dilutions using viscous or foaming liquids
    6. 6. Advanced features of the Dilution Wizard
    7. 7. How to add a wash step to a low ratio dilution
    8. 8. Aliquot Dispense Wizard
    9. 9. Semi-automatic sample preparation
    10. 10. Saving a programme as a custom method
    11. 11. Creating a Multi-Sample programme on the ML600 for the preparation of CRMs.
    12. 12. Microlab ML600 Multi-Sample dilutions for Custom solutions