Q: First, thank you for providing the wonderful resources on your website. Second, I was reviewing your Sample Preparation Guide regarding the fusing process for Titanium Oxides. I am wondering if you could clarify a few points or maybe offer some advice regarding for our case.
We are trying to measure the Ti level in vulcanized rubber samples, with the source being TiO2. We currently possess an MP-AES instrument which is not compatible with HF, so we are looking into the fusion techniques with pyrosulfate. We plan to ash the rubber sample at 850-900C so that should put the Ti in the brookite form. We then plan to fuse the residue with pyrosulfate prior to digestion and analysis on the MP-AES. I have two questions regarding these steps.
- When fusing the samples with potassium pyrosulfate, I see the ashing guide recommends 500C, but for how long do you recommend? We possess a fluxer so the heating and agitation is automated.
- The guide indicates that the fuseate is soluble in acids such as 30% sulfuric. For the MP-AES, we would prefer to be in dilute nitric acid or dilute aqua regia. Do you think the fuseate would be soluble in concentrated nitric or concentrated 1:3 nitric + HCL? Also, do you think the material will be stable once diluted to 2% nitric or dilute 1:3 nitric + HCL (~ 1% nitric & 1.6% HCL), which are the typical run conditions?